Each time QC is tested, the 24-hour clock resets.Both low and high level control must be tested every 24 hours.Control solutions must be dated with both the open and discard dates.Opened control solutions expire 3 months from opening or the “Use by” date on the bottle, whichever comes first.You may discard the code key that comes with every vial of strips.Test strips can be used until the expiration date printed on the vial.Improperly stored Test Strips will produce inaccurate results and must be discarded. There is a dessicant in the cap that keeps the strips from being exposed to humidity. Test strips are stored in the tightly capped original container.Only Accu-Chek Inform II Test Strips and Glucose Control Solutions can be used with the hospital Inform II glucose meter.Docking the meter also synchronizes the information in the meter with the system (patient lists, operator lists, strips and controls).A drained battery will take 1 hour to reach full charge.Return the Inform II meter to any base unit when finished with patient testing to charge the battery.Overview of Front of Glucose Meter Test strip port Alphanumeric touch screen Power button Quality Control – 2 vials of liquid quality control (“hi/lo”).